Sensor calibration, Sensor calibration 45 – MTS Series 793 User Manual

Page 45

Advertising
background image

About Calibration

MTS Series 793 Tuning and Calibration

Introduction

45

Sensor Calibration

Sensors convert a measured mechanical value (such as force, displacement, or
pressure) into a corresponding electrical signal. Each sensor requires
conditioning (such as AC or DC excitation) in order to output a feedback signal
that can be used by your servo controller.

Sensor/Conditioner Signal Diagram

Sensor output

Each sensor/conditioner pair must be calibrated to output a voltage that is
proportional to the measured output (which may be displacement, force, or some
other dimension).

For example, a 10-cm LVDT/AC conditioner pair (connected to a 10-cm
actuator) is typically calibrated to output:

0 volts at the piston midstroke position (0 cm)

+10 volts at maximum piston retraction (–5 cm)

–10 volts at maximum piston extension (+5 cm)

Sensor calibration

data base

The calibration procedure creates a calibration data base for each range of a
sensor. The data base that is created includes:

Calibration data points

Sensor information (model, type, serial number, calibration date)

Equipment information (identifies the equipment used in the calibration)

Conditioner information (serial number, model number, excitation voltage,
circuit parameters)

Force Sensor

Calibration

Calibrating a force sensor requires a load standard. A load standard can be a
special calibrated force sensor with its own electronics or a set of calibrated dead
weights.

Sensor

Conditioner

Module

Excitation

Signal

Sensor

Signal

Feedback

Signal

Advertising