How to Perform Advanced Tuning Techniques

MTS Series 793 Tuning and Calibration

Tuning Procedures

109

4. Set the bandwidth for the application of Limiting P Gain.

Use the Bandwidth control to define the bands in which Limiting P Gain is
applied.

5. Select the CLC control mode to test system response.

Adjust the Active P Gain if the actuator’s displacement response is
sluggish.

Adjust the Limiting P Gain if the actuator’s force response is sluggish.

6. Save the tuning values.

Standalone FlexTest SE Controllers: Select Setup > Open/Save
Parameters
> <<Save>>.

Automated Controllers: In the Station Manager window’s File menu,
select Save Parameters.

Setting the Limiting P Gain bandwidth of the limiting channel

Upper Limit

Limiting P Gain is turned on only within the upper and lower bands.
The bands are defined by the Bandwidth control, as shown in the
illustration.

Lower Limit

10 kN

0 kN

5 kN

-5 kN

-10 kN

Bandwidth 2.5 kN