MTS Series 793 Tuning and Calibration

How To Warm Up Station Hydraulics

Tuning Procedures

54

2. If necessary, turn on hydraulic pressure.

Note

The HPU can be configured for “first on”. If this is the case, start the
appropriate HSM.

A. In the power selection box, click

(Low) and then

(High)

for the HPU. If an HPU is not listed, start the HPU at the pump.

B. If an HSM is present, click

(Low) and then

(High) for the

appropriate HSM.

3. Check system interlocks

If the Hydraulic Interlock or Program Interlock indicator is lit, determine
the cause, correct it, and then click Reset.

If either interlock indicator lights again, you will need to determine the
cause of the interlock and correct it before proceeding.

4. Press Run to start the function generator.

5. Open the Tuning submenu or panel.

Standalone FlexTest SE Controllers:

Select Tuning > Manual.

Automated Controllers:

A. Click

to display the Tuning panel.

B. In the Tuning panel, click the Adjustments tab.

6. If necessary, select an access level of Tuning.

Standalone FlexTest SE Controllers: Config > Access Level > Tuning.

Automated Controllers: In the Station Manager window’s toolbar, select
an access level of Tuning.