How to Manually Tune the Control Loop
MTS Series 793 Tuning and Calibration
Tuning Procedures
81
Note
You can also use the optional Scope or an external oscilloscope to adjust
I Gain by monitoring the Error signal. For more information, see
11. Save your tuning settings.
Standalone FlexTest SE Controllers: Setup > Open/Save Parameters >
<<Save>>
Automated Controllers: On the Station Manager File menu, select Save
Parameters As.
You can either select an existing parameter set or enter the name of a new
parameter set to save.
About saving elements
of a parameter set
It is important that you save your parameter set as you complete the various
parts that make up a parameter set. Throughout this manual you will be
performing discrete procedures while building a single parameter set.
•
If you have already established your default parameter set for the
current station and you are creating a new parameter set for a specific
test, save the parameters with a different name.
•
Different tests and/or specimens may require different parameter sets.