Force sensor calibration: detailed procedure, Force sensor calibration: detailed procedure 159, Task 1, “get things ready – MTS Series 793 User Manual

Page 159

Advertising
background image

Force Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

159

Force Sensor Calibration: Detailed Procedure

Task 1

Get things ready

Perform the following before you start sensor calibration.

1. Locate relevant documentation.

You need information about the force sensor such as the serial number,
model number, excitation voltage, capacity, etc. The information can
be found on the Force Transducer Calibration Data sheet included with
a calibrated force sensor, or the Final Inspection card included with all
MTS sensors.

You need calibration identification numbers for the force standard (or
any other calibration equipment) that will be used for this calibration
procedure. The calibration information is usually on a sticker attached
to the equipment.

You need the serial number of the Digital Universal Conditioner.

You will need shunt calibration resistors.

2. Set up to monitor load standard output.

Typically, you should monitor the load standard output signal with a digital
volt meter (DVM) when you calibrate a force sensor.

3. Open a station configuration file.

You need a station configuration file that includes a control channel with a
control mode that uses the sensor you intend to calibrate.

Also, to monitor the force sensor signal with an external DVM, ensure that
you have allocated an analog output resource (readout channel) in the
Station Builder program.

On the File menu, select Open Station to open the appropriate
configuration file on the Open Station window.