Task 5, “gain/linearization calibration, Task 5 gain/linearization calibration – MTS Series 793 User Manual

Page 204

Advertising
background image

MTS Series 793 Tuning and Calibration

Extensometer Calibration

Calibration Procedures

204

Note

It is up to the user to establish whether the recorded data points meet the
required accuracy. Typically, 1% of a reading is the normal accuracy
specification.

If they do not, you must go back to recalibrate the extensometer negative
and positive outputs.

Note

The Gain/Linearization Calibration procedure that follows applies only to
controllers equipped with full-range conditioners (e.g., Model 493.25
DUC module).

Task 5

Gain/Linearization Calibration

If you are using Gain/Linearization for your calibration type, complete the
following procedure. If not, complete Task 4 Gain/Delta K Calibration on
page 201
.

Important

Using linearization data requires specific conditioner zeroing
practices. Ensure that Electrical Zero Lock on the Offset/Zero
menu is set to Locked. Adjusting electrical zero after calibration
may invalidate linearization data.

Calibration setup

1. Ensure that the extensometer is properly attached to the calibrator frame

using the associated installation drawings and calibrator fixtures.

2. Set an initial transducer excitation voltage.

Standalone FlexTest SE Controllers: Select Setup > Strain > Calibration
> Cal Type > Gain/Linearization > Excitation (peak)

Set the Excitation (peak) voltage value (typically 5- 8 volts).

Automated Controllers:

A. In the Station Setup window’s navigation pane, locate and select the

Channels resource providing the strain feedback signal.

B. In the Station Setup window, click

.

C. In the Inputs panel, click the Calibration tab.

D. Set Type to Gain/Linearization.

E.

Set Excitation (peak) voltage value (typically 5- 8 volts).

3. If the extensometer uses a zero pin, remove the zero pin now.

4. Monitor the extensometer’s amplified output (strain).

Advertising