Station Manager Detectors Window

Reset Detectors

To reset a detector that has tripped:

1. Determine the event that caused the detector to trip and fix the problem.

2. Click Reset on the Station Controls panel in the Station Manager application, or click Reset on the

Station panel of the TestSuite TW application.

If the event that caused the detector to trip is within the detector limits, the detector indicators and
any action assigned to that detector will be reset and returned to the normal (un-tripped) state.

If the event that caused the detector to trip is still true, the detector will trip and any action assigned
to that detector will occur again.

For more information on station limits, see

Access and Set Detector Settings

(p. 110).

For more information on using the Station Manager application, see MTS Series 793 Control Software.

For More Information

Access and Set Detector Settings

(p. 110)

Using Manual Command to Install the Specimen

Positioning the actuator to install or remove a specimen must be performed using the Station Manager
application. The following procedures describe how to use the Manual Command window to position an
actuator.

An optional Model 494.05 Handset can be used with the Station Manager application to provide local,
manual control of the actuator.

For more information on using the Station Manager application, see MTS Series 793 Control Software.

Manual Command Properties

This window provides manual control of channels. Use these controls to position the actuator during
specimen installation.

112 | MTS TestSuite