Modifying the finish section, Changing the maximum test runs variable – MTS TW Elite User Manual

Page 92

Advertising
background image

•

Load becomes greater than 1.00 kN (Termination Condition)

•

Specimen failure is detected

•

One of the other branches of the Parallel Branch activity (such as a limit detection branch) completes

The Brake Distance control provides control-mode braking, which slows the command to help avoid
overshoot. The brake value is subtracted from the Termination Condition value to determine where
control-mode braking starts. In this case, the Termination Condition is 1.00 kN and the Brake Distance
is set to 100 N, at 900 N the command slows to 10% of the Test Rate (0.3 mm/s), and at 990 N, command
slows to 1% of the Test Rate (0.03 mm/s).

Modifying the Finish Section

Requirements

•

The test needs to automatically generate a test report and send it to a manager via e-mail when 24
test runs have been performed.

Modifications to the Finish section

•

Change the Maximum Test Runs variable to 24.

•

Enable the Run Report activity and select the Send report in e-mail option.

Changing the Maximum Test Runs Variable

1. Click the Variables subtab (Define tab > Variables subtab).

2. Select the Maximum Test Runs variable from the list of available variables and, in the Properties Panel,

change the default value to 24.

92 | MTS TestSuite

Modify the Example Test

Advertising