MTS TW Elite User Manual

Page 89

Advertising
background image

The test needs to ramp in displacement control at 3 mm/s not to exceed a 1 kN load level.

The test needs to slow down to minimize overshoot of the ending condition.

Modifications to the Run section

Enable the Custom Message activity and add a message.

Modify the Test Rate variable from the Go To + DAQ + Detection activity.

Add a Termination Condition of 1 kN to the Go To + DAQ + Detection activity.

Add a Braking Distance of 100 N to the Go To + DAQ + Detection activity.

Enable and Configure the Custom Message Window Activity

Overview

This task enables the disabled Customer Message Window activity in the Run section and defines its
properties.

Procedure

1. Click the workflow and expand the Run section.

2. Expand the Parallel Paths activity.

3. Right-click the disabled Custom Message Window activity and select Enable.

MTS TestSuite | 89

Modify the Example Test

Advertising