GasWatch 2 Operator’s Manual

Start Up • 10

3. When the fresh air readings have been set, the LCD displays “Adj” for 2

seconds which prompts you to release the AIR button. The unit will set
the reading to 0 ppm for the GW-2C and GW-2H and to 20.9% oxygen
for the GW-2X.

4. The unit then returns to normal operation and the display indicates the

current gas concentration.

Turning Off the GasWatch 2

1. Press and hold the POWER/MODE button for about five seconds to

turn off the unit. The buzzer will pulse while the POWER/MODE button
is being pressed before the unit turns off.

2. Release the button when the LCD is blank. The unit is off.