Advertising
background image

GasWatch 2 Operator’s Manual

Displaying and Setting the Alarm Points • 18

3. As soon as segments appear on the display (approximately one

second), release the AIR button. When the unit “beeps,” release the
POWER/MODE button to put the GasWatch 2 into Alarm Point
Adjustment Mode. The LCD should display the Low Alarm setting for
the target gas and the battery level.

NOTE: If the LCD should show “CAL” in the lower left corner, the

GasWatch 2 is in Calibration Mode. You will need to press and
hold the POWER/MODE button to turn off the unit. Begin again
with step 2 above.

4. Use the POWER/MODE button to cycle through the alarm points by

pressing and releasing. If you press and hold the POWER/MODE
button, you may turn off the unit. When an alarm point is displayed, use
the AIR button to increase the alarm point. If you pass the desired
setting, continue increasing the alarm point until it reaches the
maximum setting and it will “wrap around” to the minimum setting.

5. If you adjust an alarm setting, press the POWER/MODE button to save

the new setting. The next alarm point will be displayed.

L O

Figure 3: LCD in Alarm Adjustment Mode, GW-2C & GW-2H

Alarm point

.

10.0

Battery

level

ppm

Rising
Alarm

Alarm
Name

L O

Figure 4: LCD in Alarm Adjustment Mode, GW-2X

Alarm point

.

19.5

Battery

level

%

Falling
Alarm

Alarm
Name