15 • Operation

GasWatch 2 Operator’s Manual

Resetting Gas Alarms

To reset a gas alarm, press the POWER/MODE button after the gas
reading falls below the low alarm point (or above for the GW-2X). If a TWA
or STEL alarm has been activated, it cannot be reset unless you turn off
the unit.

NOTE: Even though the gas concentration may have returned to normal

or may have fallen below the alarm point, the alarm indications will
continue until you have reset the alarm using the MODE/POWER
button.

Responding to Alarms

This section describes response to gas, over range, battery, sensor failure,
and system failure alarms.

Responding to Gas Alarms

1. Follow your established procedure for an increasing gas condition or a

decreasing oxygen condition.

2. Reset the alarm using the POWER/MODE button once the alarm

condition has been cleared.

Responding to an Over Range Alarm

WARNING: An over range condition may indicate an extreme toxic

gas or oxygen concentration. Confirm the gas
concentration with a different GasWatch 2 or with another
gas detecting device.

1. Follow your established procedure for an increasing gas condition.

2. Reset the alarm using the MODE/POWER button once the alarm

condition has cleared.

3. Calibrate the GasWatch 2 as described in the calibration section of this

manual.

4. If the over range condition continues, you may need to replace the

sensor.

5. If the over range condition continues after you have replaced the

sensor, contact RKI Instruments, Inc. for further instructions.