GasWatch 2 Operator’s Manual

Operation • 16

Responding to Battery Alarms

WARNING: The GasWatch 2 is not operational as a gas monitoring

device during a dead battery alarm. Take the GasWatch 2
to a non-hazardous area and change the battery as
described in “Changing the Lithium Battery”.

The GasWatch 2 is fully functional in a low battery warning condition.
However, only a couple of days of operation may remain depending on
certain conditions such as alarm occurrences. Change the battery as soon
as possible when a low battery warning occurs as described in “Changing
the Lithium Battery”.

NOTE: Alarms and the back light feature consume battery power and

reduce the amount of operating time remaining.

Responding to a Sensor Failure Alarm

1. Try calibrating the sensor first, as described in “Calibration”, before

replacing it.

2. If the sensor failure continues, replace the sensor as described in

“Replacing a Sensor”.

3. If the sensor failure condition continues after you have replaced the

sensor, contact RKI Instruments, Inc. for further instructions.

Responding to a System Failure Alarm

1. If a system failure occurs, try turning off the unit and turning it on again.

2. If the unit is still in system failure, contact RKI Instruments, Inc. for

further instructions.