Displaying and setting the alarm points – RKI Instruments GasWatch 2 Manual User Manual

Page 22

Advertising
background image

17 • Displaying and Setting the Alarm Points GasWatch 2 Operator’s Manual

Displaying and Setting the Alarm
Points

The GasWatch 2 allows you to display and set the alarm points. There is a
Low Alarm point and a High alarm point on all three models. The GW-2C
and GW-2H also have STEL and TWA alarm points. The alarm points and
their factory settings for the three GasWatch 2 models are summarized in
the table below:

In the table above, Low Alarm and High Alarm for the GW-2C and GW-2H
refer to a rising concentration of the target gases. The Low Alarm is
triggered at the Low Alarm concentration listed in the table. For the GW-2C
that would be 25 ppm CO, and for the GW-2H that would be 10.0 ppm. The
High Alarm is triggered when the High Alarm concentration is reached.

For the GW-2X, Low Alarm is triggered when the concentration of oxygen
falls below 19.5%. When the concentration of oxygen rises above 23.5%,
High Alarm is activated.

Display and Adjust the Alarm Points

1. Make sure the GasWatch 2 is turned off. The LCD should be blank.

2. Press and hold the AIR button, then press and hold the POWER/MODE

button.

Table 4: Alarm Points of the Four Target Gasses

Model

Low Alarm

High Alarm

STEL

TWA

GW-2C (CO)

25 ppm

50 ppm

200 ppm

25 ppm

GW-2H (H

2

S)

10.0 ppm

30.0 ppm

15.0 ppm

10.0 ppm

GW-2X (O2)

19.5%

Decreasing

23.5%

Increasing

N/A

N/A