GasWatch 2 Operator’s Manual

Maintenance • 24

Maintenance

This section describes troubleshooting procedures for the GasWatch 2. It
also describes how to change the GasWatch 2’s battery as well as how to
replace the sensor cover and gas sensor.

WARNING: RKI Instruments, Inc. recommends that service,

calibration, and repair of RKI instruments be performed
by personnel properly trained for this work. Replacing
sensors and other parts with original equipment does not
affect the intrinsic safety of the instrument.

Troubleshooting

The troubleshooting table describes error messages, symptoms, probable
causes, and recommended action for problems you may encounter with
the GasWatch 2.

Table 5: Troubleshooting the GasWatch 2

Symptoms

Probable

Causes

Recommended Action

• The LCD is

blank.

• The unit may

have been
turned off.

• The battery

may need to
be replaced.

1. To turn on the unit, press

and hold the MODE/
POWER button.

2. If the unit does not turn on,

replace the battery.

3. If the difficulties continue,

contact RKI Instruments,
Inc. for further instruction.

• The LCD shows

abnormally high
or low readings
but other gas
detection
instruments do
not.

• The unit may

need to be
recalibrated.

• The sensor

may need
replacement.

1. Recalibrate the unit.
2. Replace the sensor and

calibrate the unit.

3. If the difficulties continue,

contact RKI Instruments for
further instruction.