25 • Maintenance

GasWatch 2 Operator’s Manual

Replacing the Lithium Battery

WARNING: Replace the lithium battery in a non-hazardous

environment.

Replace the lithium battery when the battery icon indicates
that the unit is in low battery warning. When in low battery
warning, only one battery level indication bar remains and it
is flashing.

NOTE: Use a CR 2450 battery manufactured by Sony, Eveready, Maxell

Hitachi, or Toshiba to maintain the CSA classification of the
GasWatch 2.

• “FAIL” displays

during span or
zero adjustment.

• The calibration

value may not
match the
cylinder gas
concentration.

• The sample

gas is not
reaching the
sensor
because of a
bad
connection.

• The calibration

cylinder may
be out of gas
or is outdated.

• The sensor

may need
replacement.

1. Check all calibration tubing

for leaks or for any bad
connections.

2. Make sure the GasWatch 2

has been properly set up
for calibration.

3. Verify that the calibration

cylinder contains an
adequate supply of fresh
test sample.

4. If the fail condition

continues, replace the
sensor.

5. If the difficulties continue,

contact RKI Instruments,
Inc. for further instruction.

• “FAIL SYS” is

indicated on the
LCD

• A microproces-

sor failure has
occurred.

1. Turn off the unit and turn it

on again.

2. If difficulties continue,

contact RKI Instruments,
Inc.

Table 5: Troubleshooting the GasWatch 2

Symptoms

Probable

Causes

Recommended Action