Appendix a: mri applications – RKI Instruments GasWatch 2 Manual User Manual

Page 37

Advertising
background image

GasWatch 2 Operator’s Manual

Appendix A: MRI Applications • 32

Appendix A: MRI Applications

When the GasWatch 2 is used in MRI applications or applications where a
strong magnetic field is present on a regular basis, the magnetic field will
cause damage to the instruments’s buzzer over time. This damage may
eventually render the buzzer inoperable.

WARNING: In MRI or similar applications, do not rely on the buzzer

for alarm indications. The GasWatch 2 also uses a
vibrator and visual indications for alarms. Make sure you
are using these indications when using the GasWatch 2
for this type of application.

The damage caused to the buzzer by a magnetic field in these applications
is not covered under warranty.

81-GW2H-LV

Calibration kit for GW-2H, one 34 liter aluminum
gas cylinder (25 ppm H

2

S in nitrogen), regulator,

calibration cup, case, & tubing

81-GW2X-LV

Calibration kit for GW-2Xone 34 liter steel gas
cylinder (100% nitrogen), regulator, calibration
cup, case, tubing

82-0001RK

Spring bar tool

ES-1821

Carbon monoxide sensor

ES-1821L

Carbon monoxide sensor, for low humidity use

ES-1827

Hydrogen sulfide sensor

ES-1827L

Hydrogen sulfide sensor, for low humidity use

OS-BM2

Oxygen sensor

Table 6: Parts List