Chapter 3 setup menus, Basic setup menu, Chapter 4 setup menus – SATEC C191HM User Manual

Page 29: 1 basic setup menu

Advertising
background image

Chapter 4 Setup Menus

20

Chapter 4 Setup Menus

4. CHAPTER 3 SETUP MENUS

NOTE: Instrument setup can be performed directly on the front panel using the setup
menus or via communications using PComTest communication software. PComTest is
supplied with your instrument and provides full setup capabilities for your instrument. For
information on using PComTest, refer to the user documentation supplied with your
instrument.

Setup Display PComTest
Basic

+

+

Communication port

++

+

User Selectable options

++

+

Analog output

+

+

Digital inputs

+

+

++ Recommended method

Alarm/Event set points

+

+

Pulsing output

+

+

Pulse counter

+

+

Assignable registers

-

++

Display

++

-

4.1 Basic Setup Menu

This menu contains the basic configuration options which define the general
operating characteristics of your instrument, such as wiring mode, input scales,
the size of the RMS averaging buffer, etc. Table 4-1 lists the basic setup options,
their code names and applicable ranges.

Activate the middle window to scroll through the list of available options, and
then activate the lower window to set the option value.

To select and view a setup option:

Press to activate the middle window
Use the up/down arrow keys to scroll to the desired option. The current

value for this option appears in the lower window.

To change the value of the selected option:

Press to make the lower window active.
Press the up/down arrow keys to scroll to the desired value.
Press to store the selected value, or press to quit the

setup menu.

SELECT

SELECT

bASc

4L-n

ConF

ENTER

ESC

SELECT

CHG

ENTER

bASc

ENTER

Advertising