Chapter 2 Balancing Extension Module

Detection

Selects the type of signal detection and scaling for the
measurement:

Peak - The data collector measures the dynamic signal

from zero to peak voltage. Use this or the RMS setting for
most acceleration and velocity measurements.

Pk-Pk - The data collector measures the dynamic signal

from the minimum peak to the maximum peak. Use this
setting for most displacement measurements.

RMS - The data collector measures the dynamic signal as

the square root of the mean of the square. Use this or Peak
for most acceleration and velocity measurements. Also use
this for voltage measurements.

Peak: Scaled from RMS as

2

RMS

Pk-Pk: Scaled from RMS as 2

2

RMS

RMS: The Root Mean Squared overall
calculated from the FFT or time data.

Defaults to Peak for Accel

Defaults to Vel and Pk-Pk for Disp.

No. Averages

Enter the number of averages (1 to 8) for the measurement.

4 (default)

Solution

Select the balancing solution. The parameter is available only
when you are two-plane balancing:

Dynamic - The correction weights will be calculated for

both planes 1 and 2.

Static-Couple - The correction weights will be calculated

as a combination of Static and Couple. The Couple balance
weights will be presented as the same weight in both
planes 1 and 2, with a phase angle of 180

°

difference. The

Static part of the solution will be calculated at the center
between planes 1 and 2.

Dynamic (default)

Static-Couple

Table 2 - Balance Extension Module Parameters

Parameter Description

Value