Balancing Extension Module Chapter 2
5.
When the speed, vibration, and phase values are stable, press Enter.
The trial weight measurement is complete and the results are stored in the
Vibration Summary Table.
A message appears asking if the trial weight is still attached:
• Press F2 (Yes) if the weight is still attached to the rotor.
• Press F3 (No) if the weight has been removed from the rotor.
6.
Shut down the machine.
TIP
It is good practice to remove the trial weight.
TIP
Anytime during the initial run, you can do the following:
•
Press F4 (Esc) to exit the balancing program.
•
Press 0 (Shift) and F1 (Go to) simultaneously to go back to a
previous step in the balance run.
Move Around in a Balance Run on page 60
for details.
•
Press F2 (Summary) to display the results in the Vibration
Summary Table.
Review Balancing Measurements on page 63
for details.