You are prompted to verify the transducer is attached to the bearing
nearest Plane 2.
5.
Verify the transducer and Plane 2 and press OK.
The data collector initiates the initial run measurements for Plane 2. The
Initial Run – Plane 2 screen displays the measurement results.
6.
From the Initial Run – Plane 2 screen, press Enter.
7.
Attach a Trial Weight to Plane 1.
The Add Trial Weight – Plane 1 screen appears.
8.
Stop the machine.
9.
On the data collector, enter the Mass, Angle and Radius for the trial weight
you are attaching to Plane 1.
• A negative number can be entered if trial weight is removed, instead of
added.
• Trial weights are typically attached at 0°.
• Unless you wish the data collector to estimate a trial weight, the Radius
entry is not necessary, so long as you position the trial, correction, and
trim weights at the same radius.
Add the Trial Weight on page 28
.