761 SD Compact IC / Instructions for Use 8.

761.1043

100

6.2.7

Cleaning the suppressor

It may be necessary to clean the suppressor in the following cases:

•

Increased counter-pressure on the suppressor connec-

tion tubing

•

Irremediable blockage of the suppressor

(Solutions can no longer be pumped through the sup-

pressor)

•

Irremediable obstruction of the suppressor

(The suppressor can no longer be switched to the next

position)

Proceed as follows (see Figure 18) to clean the connector and the sup-

pressor rotor:

1 Disconnect the suppressor from the IC system

•

Disconnect the suppressor inlet capillary ("Eluent") 37 from the

separating column 36.

•

Disconnect the suppressor outlet capillary ("Detector") 38 from

the "inlet capillary to detector block" 39.

•

Disconnect the inlet capillaries ("H2O") 42 and ("H2SO4") 45

from the filter units 56.

2 Dismantle the suppressor

•

Unscrew the nut 89 from the suppressor holder 92.

•

Pull out the connector 90 and suppressor rotor 91 from the

suppressor holder 92 (the connector and rotor normally stick

together).

•

Detach the connector 90 from the suppressor rotor 91.

3 Clean the inlet and outlet capillaries

•

Connect each of the 6 capillary tubings attached to connector

90 to the high-pressure pump one after the other and pump

through ultra-pure water.

•

Check whether the solution emerges from the connector 90. If

one of the inlet or outlet capillaries remains blocked, replace

the connector 90 (order number 6.2832.010).

4 Clean the suppressor rotor

•

Clean the sealing surface of the suppressor rotor 91 using a

lint-free cloth and ethanol.