Connection sample vessel, Pump tubing, Injection valve – Metrohm 761 SD Compact IC User Manual

Page 32: Waste, Figure 10, Pump tubing for sample path, Peek coupling 6.2744.030, Pump tubing 6.1826.110, Peek coupling 6.2744.160, Fit the coupling 63

Advertising
background image

2 Installation

761 SD Compact IC / Instructions for Use 8.

761.1043

24

Figure 10: Pump tubing for sample path

57 Tubing cartridge

58 Pressure lever

59 Snap-action lever

62 Stopper orange-yellow

63 PEEK coupling 6.2744.030

64 Pump tubing 6.1826.110 for sample

65 PEEK coupling 6.2744.160

with tubing security device

2.8.2

Connection sample vessel

→

pump tubing

→

injection valve

→

waste

Proceed as follows in order to make the connections for the sample

path:

1 Connection Sample vessel - pump tubing

The sample is aspirated with the aspirating tubing 3:

•

The aspirating tubing 3 is routed through the "feedthrough for

aspirating tubing" 2 out of the front door into the sample ves-

sel. It is fixed in position on the feedthrough 2 with a rotary

nipple (see Section 2.9.2 if using the 766 IC Sample Proces-

sor).

•

Detach the other end of the aspirating tubing 3 with pressure

screw 50 from the injection valve (it is fitted there only for de-

livery purposes) and screw it firmly onto the PEEK coupling

63 on the left-hand end of the pump tubing (flow inlet) (see

Figure 11).

65

64

62

59

63

57

58

62