6.3 Faults

and

malfunctions

6.3.1 Error

messages

If any type of malfunction occurs during operation of the 761 SD Com-

pact IC, this is shown by error messages in the PC program, which ap-

pear either in an error window or in the

SYSTEM STATE

window.

Follow the instructions listed in the Error window and close this win-

dow with

<OK>

.

You fill find further details of the error messages of the

SYSTEM STATE

window, their possible causes and the procedure for rectifying them in

the Software Instructions for Use «IC Net 2.3» Section 4.5.

6.3.2

Malfunctions and their rectification

If difficulties appear with The 761 SD Compact IC during analyses, their

causes are best investigated in The order separating column

→

high-

pressure pump

→

eluent

→

connections. Several of The malfunc-

tions which may appear are listed in The following table with details of

possible causes and countermeasures.

Malfunction

Cause

Rectification

Baseline with high
noise level, pulsa-
tion

•

Contaminated pump
valves

•

Defective piston seals

•

Clean the valves (see Sec-
tion 6.2.5
)

•

Replace the piston seals
(see Section 6.2.5)

Drift of the
baseline

•

Thermal equilibrium not
yet reached

•

Leakage in the system

•

Condition system with
heating switched on

•

Check connections and
seal them

Considerable
pressure drop

•

Leakage in the system

•

Check connections and
seal them

Considerable
pressure rise

•

Blockage of filter unit
PEEK 6.2821.120

•

Change of column pack-
ing by injection of con-
taminated samples

•

Replace the filter
6.2821.130 (see Section
2.3.3
)

•

Regenerate the column or
replace the column (see
Section 6.2.4)

Note:
Samples should always
be microfiltered

Chromatograms
with poor resolu-
tion, Change in
the retention times

•

Deterioration in separa-
tion efficiency of the IC
column

•

Regenerate the column or
replace the column (see
Section 6.2.4)

Extreme peak
broadening,
splitting (double
peaks)

•

Dead volume at the col-
umn ends

•

Regenerate the column
(see Section 6.2.4)