Metrohm 761 SD Compact IC User Manual

Page 44

Advertising
background image

2 Installation

761 SD Compact IC / Instructions for Use 8.

761.1043

36

1 Detach the column connection capillary 33 from the

PEEK coupling

•

Undo the pressure screw 50 with which the column connec-

tion capillary 33 is screwed to the PEEK coupling 54 (transi-

tion to the suppressor). Take this end of the capillary and

route it into a beaker.

The column connection capillary 33 should now connect the in-

jection valve to the beaker.

2 Open and connect the system

•

Open the

startup.smt

system, connect it and make the same

settings as described in Section 2.13.1.

3 Rinse the pulsation dampener

•

Click on the

<On>

button for

IC pump

to switch on the high-

pressure pump and rinse the pulsation dampener 30 filled

with isopropanol for approx. 10 minutes with eluent.

•

Click on the

<Off>

button for

IC pump

to switch the high-

pressure pump back off again.

4 Connect the column connection capillary 33 back to

the PEEK coupling

•

Take the end of the column connection capillary 33 out of the

beaker and screw it back onto the PEEK coupling 54 (as a

transition to the suppressor) with a pressure screw 50.