Metrohm 761 SD Compact IC User Manual
Page 44

2 Installation
761 SD Compact IC / Instructions for Use 8.
761.1043
36
1 Detach the column connection capillary 33 from the
PEEK coupling
•
Undo the pressure screw 50 with which the column connec-
tion capillary 33 is screwed to the PEEK coupling 54 (transi-
tion to the suppressor). Take this end of the capillary and
route it into a beaker.
The column connection capillary 33 should now connect the in-
jection valve to the beaker.
2 Open and connect the system
•
Open the
startup.smt
system, connect it and make the same
settings as described in Section 2.13.1.
3 Rinse the pulsation dampener
•
Click on the
<On>
button for
IC pump
to switch on the high-
pressure pump and rinse the pulsation dampener 30 filled
with isopropanol for approx. 10 minutes with eluent.
•
Click on the
<Off>
button for
IC pump
to switch the high-
pressure pump back off again.
4 Connect the column connection capillary 33 back to
the PEEK coupling
•
Take the end of the column connection capillary 33 out of the
beaker and screw it back onto the PEEK coupling 54 (as a
transition to the suppressor) with a pressure screw 50.