Metrohm 761 SD Compact IC User Manual

Page 87

Advertising
background image

5.1 Operation with manual sample change

761 SD Compact IC / Instructions for Use 8.

761.1043

79

2 Calibration

Dry off the free end of the sample aspirating tubing 3 with a

clean cloth.

Then immerse it in the lowest MCAL standard (mg/L H

3

PO

4

plus basic content of nitrate and sulfate). The MCAL standard

bottles are supplied as accessories (6.2321.000).

Start the system by selecting

SYSTEM /

Control

/ Start determi-

nation

(see Software Instructions for Use «IC Net 2.3», Sections

4.3.3 and 4.3.4). After start, you will see the

EDIT SAMPLE

DESCRIPTION: SYSTEMS

window. You should make the fol-

lowing entries:

What is important above all is to set menu option

Calibration

level

to 1.

After completing this measurement, you should determine the

other 5 MCAL standards one after the other (250, 350, 450,
550 and 650 mg/L H

3

PO

4

plus basic content of nitrate and

sulfate) and, when doing this, increase menu item

Calibration

level

in window

EDIT SAMPLE DESCRIPTION: SYSTEMS

in-

crementally by one (this means that the measurement with
the 650 mg/L standard would then be

Calibration level

6).

In order to check calibration, you should also determine the

MCAL QC check standard with 545 mg/L H

3

PO

4

plus basic

content of nitrate and sulfate at the end of the series of cali-
bration measurements.

Detailed information on calibration can be found in the Soft-

ware Instructions for Use «IC Net 2.3», Section 7.5.