Connection of the 813 compact autosampler – Metrohm 844 UV/VIS Compact IC User Manual

Page 68

Advertising
background image

Table of contents

5 Setting up the 844/838 system

• Select

File / Open / System

in the main window of the «IC Net »

software. Open the system for the 844 UV/VIS Compact IC

that you have created in Section 2.5.4.

• Add the 838 Advanced Sample Processor to the system (via

Setup / New devices / Install new devices..

, see «IC Net»

Instructions for Use, Section 4.4.4):

•

6 Creating a program for the 838

• Open the window with the system settings of the 838 (with a

double-click on the 838 icon or via the right-hand mouse key

open

).

• Click on the

Program

tab. The 838 Advanced Sample Pro-

cessor is controlled by a command sequence on the

Program

tab.

• Create a program (see «IC Net» Instructions for Use, Section

6.26.3):

•

•

At the command Scan Wait1 the 838 will wait for an

incoming Remote signal on line 3 (defined in the subtab

Confi-

guration / Scan

). It will receive this Remote signal from a Remote

command included in the time program of the 844 (see below).

The 844 and 838 are synchronized in this way.
•

For more detailed information about the control of the 838

Advanced Sample Processors via «IC Net» please refer to «IC

Advertising