Metrohm 844 UV/VIS Compact IC User Manual
Page 85

3.3 Settings
844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051
73
Flow correction
Factor for correction of the difference bet-
ween displayed and actual flow rate of
the high-pressure pump.
Range:
0.9
...
1.25
The correction fac
tor is determined as
follows by measuring the actual flow rate
using a graduated cylinder:
Flow correction
=
Displayed flow rate
Measured flow rate
The product of parameter
Flow correction
and
parameter
Flow
(on the tab
Controls
) must not
exceed 2.75 mL/min.
Ch1/Ch2 threshold
Limit for measuring channel ratio
Ch1/Ch2.
Threshold, mAU
As long as the absorption value(s)
from channel 1, channel 2 or both
lie beneath the defined
Threshold
(limit)
the value for the Ch1/Ch2 ratio will not
be calculated.
Analog Outputs
Settings for analog signal output.
Range, mAU
The defined value is equivalent to 1 V
for the analog signal output.
Entry range:
0.1
...
2000.0 mAU
Detector settings
On the tab
Detector settings
measuring settings are defined.
Detector settings
Settings for UV/VIS detector 43/61.
VIS lamp
Display of the voltage of the VIS lamp.