the openings of the 844 UV/VIS Compact IC and use a

6.2744.010 PEEK Compression fitting to screw it onto

injection valve 46/60..

3 Tubing connection Injection valve – Waste

• Loosen connection capillary 33 from injection valve 46/60.

(sample outlet).

• Cut a piece of 6.1831.060 PTFE capillary (sample changer

accessory) to the required length, attach it to injection valve

46/60. using a 6.2744.010 PEEK compression fitting and

place the other end in a waste container.

4 Setting a method on the 813

• Use the 813 keyboard on the 813 Compact Autosampler to

set the required method (see 813 Instructions for Use).

5 Creating a time program for the 844

• In the main window of the «IC Net» software select

File / Open /

System

. Open the system for the 844 UV/VIS Compact IC that

you have created in Section 2.5.4..

• Open the window with the system settings of the 844 UV/VIS

Compact IC (with a double-click on the 844 icon or via the

right-hand mouse key

open

).

• Open the

Program

tab.

• Create a time program:

•

•

The Remote command sends a signal to the 813 via

Remote line 3. The function of this signal depends on the

method selected at the 813 (see 813 Instructions for Use,

Section 3.4).
•

The two Valve commands switch the injection valve of the

844 to the positions Fill and Inject.
•

For more detailed information about the control please refer

to Section 3.3.1 or the «IC Net» Software Instructions for Use.