Metrohm 844 UV/VIS Compact IC User Manual
Page 83

3.3 Settings
844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051
71
Event
Events for automatic remote signal output:
Inject
Switching of the injection valve to the "INJECT"
position.
Fill
Switching of the injection valve to the "FILL"
position.
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.
Hardware error
Hardware error detected at the 844 UV/VIS
Compact IC (high-pressure pump, injection
valve, or peristaltic pump not working correctly).
Alarm leak detector
Internal
leak.
PowerOn values
Output
lines
1… 8 will be set to the values
defined here after the instrument has been
switched on or after an emergency stop with
Shutdown hardware
.
Selection:
0
,
1
Pulse length
Length of a pulse in ms.
Alarm stops
The
Alarm stops
tab of the hardware settings window defines the events
for which the instrument is stopped immediately. At an alarm stop, high-
pressure pump and peristaltic pump are stopped immediately, the
running determination and the active sample queue are also stopped.