Metrohm 844 UV/VIS Compact IC User Manual
Page 81

3.3 Settings
844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051
69
Name (1
st
column)
User-definable name of the remote command
(e.g.
Start_838
).
Remote command (2
nd
column)
Setting the remote output lines 1…8.
Selection:
0
(line off, inactive, open)
1
(line on, active, 0 V)
p
(pulse)
*
(leave line in current status)
For entry of the first value, enter
1
,
0
,
p
or
*
. For
entry of the other values, move the cursor in
front of the value to be changed and enter
1
,
0
,
p
or
*
.
<Add>
Add new remote command.
<Delete>
Delete selected remote command.
<Activate>
Activate the defined remote commands for
insertion into the time program.
Configuration
The
Configuration
tab in the system settings window contains
configuration settings for the 844 UV/VIS Compact IC.
Startup with hardware
The activated devices will be started with
Startup
hardware
or at the start of a determination.
UV lamp
If this option is switched off the UV lamp will not
be switched on at
Startup hardware
or at the start
of a determination.
VIS lamp
If this option is switched off the VIS lamp will not
be switched on at
Startup hardware
or at the start
of a determination.
Column oven
If this option is switched off the column heating
will not be switched on at
Startup hardware
or at
the start of a determination.