Metrohm 844 UV/VIS Compact IC User Manual

Page 76

Advertising
background image

Table of contents

Pump pressure, MPa

Displays the currently measured pressure of the

high-pressure pump (units depend on the

settings made under

IC NET /

Global

preferences

).

Valve

Live display of the position of the injection

valve.

Remote lines

Live display of the status of the 8 remote lines.

Column temp., °C

Live display of the temperature in the column

oven.

Controls

On the subtab

Controls

injection valve 46/60., peristaltic pump 49, high-

pressure pump

27/54, column heating 115 and remote lines can be

manually controlled. In addition, the parameters for the high-pressure
pump and column heating temperature can also be set:

Valve

Injection

valve

<Inject>

Switch injection valve to "Inject" position.

<Fill>

Switch injection valve to "Fill" position.

Peristaltic pump

Peristaltic

pump

<On>

Start peristaltic pump (only enabled, with

versions 2.844.0020 or 2.844.120).

<Off>

Stop peristaltic pump (only enabled, with

versions 2.844.0020 or 2.844.120).

IC pump

High pressure pump

<On>

Start high pressure pump.

<Off>

Stop high pressure pump.

Flow, mL/min

Set value for flow rate of IC pump.

Entry range:

0.20

...

2.50 mL/min