9.1 Karl Fischer titration

■■■■■■■■■■■■■■■■■■■■■■

90

■■■■■■■■

899 Coulometer

Problem

Cause

Remedy

The Control parameters
are not set correctly.

■

Reset the Control parameters to default
values.

The titration time is
too long.

The drift during condition-
ing is not stable.

■

Wait before starting the titration until the
drift has become stable.

The parameter Start drift
is set too high.

■

Set the Start drift lower.

■

Adjust the Stop drift or use the Relative
stop drift
.

The parameter Stop drift
is set too low.

■

Set the Stop drift higher or use the Rela-
tive stop drift
.

The water quantity is too
high.

■

Observe the sample size and water con-
tents in accordance with Table "Recom-
mended sample sizes", page 31
.

The Control parameters
are set incorrectly.

■

Set the parameter Dynamics lower and
the parameter Max. rate higher.

The result is too
high.

The coulometer cell is not
yet correctly conditioned.

■

Shake the coulometer cell and wait until
the drift is stable.

The sample contains oxi-
dizable substances.

■

Possibly use a KF oven.

■

See technical literature.

The parameter Stop drift
is set too low.

■

Set the Stop drift higher or use the Rela-
tive stop drift
.

The drift correction is too
small (e.g. with the use of
a KF oven or with manual
drift correction).

■

Start the titration as long as the gas flow
still exists between the conditioning vial
and the coulometer cell.

■

Set the parameter Drift correction to
auto (= automatic drift correction).

The result is too
low.

The sample releases iodine.

■

Possibly use a KF oven.

■

See technical literature.

The parameter Stop drift
is set too high.

■

Set the Stop drift lower.

The drift correction is too
large (e.g. with excessively
high start drift, with fluctu-
ating drift or with manual
drift correction).

■

Ensure stable and lower drift.

■

Set the parameter Drift correction to
auto (= automatic drift correction).