10 Appendix

899 Coulometer

■■■■■■■■

97

10.5

Method examples for working with the Karl Fischer
oven

You will find the recommended parameter settings for working with a Karl
Fischer oven in the following table. The methods can be used either with a
sample changer with oven module (e.g. 885 Compact Oven SC) or with a
Karl Fischer oven (e.g. 860 KF Thermoprep). The table contains only para-
meter settings that differ from the default values.

Parameters

Blank value
determination
(method tem-
plate Blank)

Water content
determination
(method tem-
plate KFC-Blank)

>Conditioning

Start drift

10 µg/min *)

10 µg/min *)

Stabilizing time

30 s

30 s

>Start conditions

Request sample size

off

off

Hold at request

off

off

>Titration parameters

Extraction time

180 s

180 s

Stirring rate

12

12

>Statistics

Statistics

on

Number of determina-
tions

3

>Reports

Results

on

on

Calculations/Statistics

on

*) If you work in an environment with high humidity, then we recommend
that the parameter Start drift be set to 20 µg/min.