7 remote interface, 1 pin assignment of the remote interface, Pin assignment of the remote interface – Metrohm 899 Coulometer User Manual
Page 109: Figure 27, Pin assignment of remote socket and remote plug

■■■■■■■■■■■■■■■■■■■■■■
10 Appendix
899 Coulometer
■■■■■■■■
99
3 Confirming the initialization
Note
If the request is not confirmed within 8 s, then the procedure will
be canceled.
■
Press [BACK] twice.
Initialization is started. The process takes approximately 80 s. The
instrument will be automatically restarted after successful initializa-
tion.
10.7
Remote interface
10.7.1
Pin assignment of the remote interface
1
5
6
9
5
1
9
6
Figure 27
Pin assignment of remote socket and remote plug
The above figure of the pin assignment applies for all Metrohm instru-
ments with 9-pin D-Sub remote connector.
Table 8
Inputs and outputs of the remote interface
Pin No.
Assignment
Function
1
Output 0
Ready/EOD
2
Output 1
-
3
Output 2
Titration
4
Output 3
Cond OK
5
Output 4
Error
6
0 volt (GND)
7
+5 volts
8
Input 0
Start
9
Input 1
Stop