Figure 6, Mounting the addition and aspiration tube – Metrohm 899 Coulometer User Manual
Page 23

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
899 Coulometer
■■■■■■■■
13
3.2.3
Mounting the addition and aspiration tube (utilization with Ti
Stand)
6.1437.000
6.2713.000
6.2713.000
6.1439.010
1
2
4
5
6
3
Figure 6
Mounting the addition and aspiration tube
Insert the addition and aspiration tube as follows into the coulometer cell:
1 Attach the 6.2713.000 ground-joint sleeve that has been cut to size
to the ground joint of the 6.1437.000 stopper.
2 Insert the stopper into the 6.1439.010 addition and aspiration tube.
3 Attach the 6.2713.000 ground-joint sleeve that has been cut to size
to the ground joint of the addition and aspiration tube.
4 Insert everything together into the ground-joint opening.
5 Connect the tubing for the reagent addition at the upper connector
of the addition and aspiration tube (5).