5 control, Control – Metrohm 899 Coulometer User Manual
Page 53

■■■■■■■■■■■■■■■■■■■■■■
5 Operation
899 Coulometer
■■■■■■■■
43
5.5
Control
Menu
▶ Control
In the dialog Control, the settings for the execution of a single determi-
nation or of one sample series are defined.
Sample table
If this parameter is activated, the sample data for a sample series can be
entered in a table (see Chapter 5.7, page 45).
Selection
on | off
Default value
off
Autostart
If this parameter is activated, a new determination is started automatically
at the end of a determination. This continues until the number specified
has been reached (see Number of autostarts).
Selection
on | off
Default value
off
Note
If you are working with a sample changer with oven module (e.g. 885
Compact Oven SC), then the Autostart parameter must be set to off,
because the sample changer starts the titration.
Number of autostarts
This parameter is visible only when Autostart = on.
Number of automatic starts.
Input range
1 ... 50
Selection
table
Default value
table
table
The number of automatic starts corresponds to the number of samples
in the sample table.