4 connecting sensors, 1 connecting a generator electrode, Connecting sensors – Metrohm 899 Coulometer User Manual

Page 27: Connecting a generator electrode, Figure 9, Unscrewing the cover from the generator electrode

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

899 Coulometer

■■■■■■■■

17

3.4

Connecting sensors

The measuring interface contains the following measuring inputs:

Gen. for a generator electrode

Ind. for a double Pt electrode

Temp. for a temperature sensor of the Pt1000 type

Caution

Under all circumstances, avoid mixing up the electrode cable from the
indicator electrode with the one from the generator electrode. Attach
corresponding markings on the screw heads of the cables.

3.4.1

Connecting a generator electrode

Screwing the electrode cable to the general electrode

1 Unscrew the cover of the generator electrode.

1

Figure 9

Unscrewing the cover from the generator electrode

2 Screw the 6.2104.120 electrode cable tightly onto the generator

electrode.