Metrohm 899 Coulometer User Manual
Page 79

■■■■■■■■■■■■■■■■■■■■■■
6 System settings
899 Coulometer
■■■■■■■■
69
Selection
English US | French FR | German CH | German
DE | Spanish ES
Default value
English US
Balance
If you have connected a balance, then you must define the balance type
here.
Selection
AND | Mettler | Mettler AT | Mettler AX |
Ohaus | Precisa | Sartorius | Shimadzu
Default value
Sartorius
The following table indicates the balance type that needs to be selected
for the balance model:
Balance
Balance type
AND
AND
Mettler AB, AE, AG, AM,
AJ, PE, PM, PJ, PR, XP, XS
Mettler
Mettler AT
Mettler AT
Mettler AX, MX, UMX, PG,
AB-S, PB-S
Mettler AX
Ohaus Voyager, Explorer,
Analytical Plus
Ohaus
Precisa
Precisa
Sartorius
Sartorius
Shimadzu BX, BW
Shimadzu
Editing the COM1 settings
Menu
▶ System ▶ External devices ▶ COM1 settings
The interface parameters for the connected balance are set under COM1
settings.
Baud rate
Transfer rate in characters per second.
Selection
1200 | 2400 | 4800 | 9600 | 19200 | 38400 |
57600 | 115200
Default value
9600
Data bits
Number of data bits.