11. Plot the results as a function of sample concentration (Figure 2.7). The overall trend

should be linear.

Fig. 2.7. Plot of concentration calculated by Experion software as a function of expected
concentration
.

Expected Performance of Test 2. Based on results obtained using a total of 6 chips.

Measurement

Typical Results

Quantitation reproducibility

%CV £ 20% for total RNA (25–500 ng/µl)

If your results do not appear as they do in these examples, refer to the Troubleshooting
section of the Experion RNA StdSens and Experion system instruction manuals or contact
Bio-Rad technical support. Typically, problems with reproducibility are due to problems with
pipetting or chip loading. For tips on optimizing your results, refer to the Experion Training
Video: Chip Loading and the Essential Practices and Troubleshooting sections of the
Experion system instruction manual.