8

Detailed Instructions

The following pages provide detailed instructions for each step of the assay
procedure, including sample preparation, running the assay, and reading
the plate with Bio-Plex Manager

and Luminex xPONENT software.

1. Prepare the Samples

Considerations

n

The degree of phosphorylation of a given analyte is highly dependent

on the cell type and cell stimulation or treatment conditions

n

Cell lines may vary in their signaling responses to the same stimulation

n

The suggested final protein concentration range in the assay is

3–200 μg/ml (0.15–10 μg per assay well) except for Pl3K p85 (Tyr

458

)

which is 31–1,000 μg/ml (1.6–50 µg/well)

n

Optimization of cell lysate concentration may be needed based on

target protein expression levels

n

Cell lysate should be clear of particulate matter before use

Cell Lysates

The Bio-Plex Pro

cell signaling reagent kit (catalog #171-304006M)

is required for preparing lysates derived from cell culture and tissue
samples. Just before use, prepare an adequate volume of cell lysis buffer
by adding PMSF and cell lysis factor QG.

n

Prepare 500 mM PMSF by dissolving 0.436 g PMSF in 5 ml DMSO.
(Store as aliquots at –20°C). Add PMSF to the cell lysis buffer at a final
concentration of 2 mM

n

Reconstitute cell lysis factor QG to 100x with 250 μl of diH

2

O and vortex

to mix. Add the reconstituted factor to the cell lysis buffer to a final 1x
working concentration