Bio-Rad Bio-Plex Pro™ Magnetic Cell Signaling Assays User Manual

Page 31

Advertising
background image

29

Table 13. Troubleshooting guide (continued).

Problem

Possible Causes

Possible Solutions

Low bead count

Cell debris in lysate

Centrifuge at 15,000 x g for 10 min

not cleared

at 4°C to remove cellular debris

Assay plate not shaken

Shake plate at 900–1,100 rpm

enough prior to reading

for 30 sec before data acquisition

Clogged reader

Refer to the troubleshooting

guide in the Bio-Plex system

hardware instruction manual

(bulletin 10005042)

Miscalculation of bead dilution

Check the calculations and be

careful to add the correct volumes

Clumping of stock beads

Vortex the stock vial at different

in vials

angles for 30 sec at medium speed

before aliquoting beads

Vacuum setting too high

Adjust pressure to –1 to –3" Hg.

during suction of assay plate

Generally, 100 μl liquid should

take 3–4 sec to clear from the well

Reader needle height

Adjust the needle height to coincide

incorrect

with the plate type provided in the kit

Beads lost

Use recommended magnetic
washer settings with correct
soaking time

continues

Advertising