a.

An SV-3 solenoid valve is used with the Model 2110 fraction
collector.

b.

Plug the SV-3 valve directly into the diverter valve receptacle on
the back of the BioLogic LP instrument.

Section 9
Power Connections

1. Plug the line power cords into the rear panel of the BioLogic LP and

the fraction collector.

2. Power for the 1327 chart recorder is supplied from the 12 volt outlet on

the rear panel of the BioLogic LP (next to the power connection for the
UV optics module). Locate the thin black power cable, and connect
this between the BioLogic LP and the power inlet for the 1327 (next to
the DIN socket, chart recorder rear panel). The 1327 chart recorder
may also be powered by a separate AC adaptor.

Section 10
Pump Setup

1. Adjust platen pressure screw (on pump head) as shown in Figure 4.

2. Using a screwdriver or coin, turn the screw counterclockwise as far as

it will go, then turn clockwise three full turns.

Fig. 4.

1

Alarm

2

3

4

5

6

7

8

0

9

C

.

Platen Adjust
Screw

Orange Lock-Ring

1.6 mm ID Tubing

Female
Luer Fitting

See Detail

7