Bio-Rad BioLogic LP System User Manual

Page 15

Advertising
background image

b.

Press the soft key for CALIBRATE.

c.

Enter the Conductivity Cell Constant displayed on the conductivity
flow cell cable.

2. In Manual mode:

b.

Press the COND soft key.

c.

Select Min/Max.

d.

In the Min field, enter the conductivity of “A” buffer (determined
above in Step 6 of Section 13).

e.

Press SETMAX.

f.

Enter the conductivity of “B” buffer (determined above in Step 5 of
Section 13).

g.

Press OK.

Section 15
Calibrating the Pump

You must tell the system which tubing is installed in the pump head.

In Manual mode:

1. Press the PUMP soft key.

2. Select Flow, then Calibrate.

3. Select the tubing size you are using (1.6 mm ID).

4. Select Nominal.

5. The pump is now calibrated, using the Nominal flow rate factor for

1.6 mm ID tubing.

Section 16
Setting Up Your Fraction Collector

If you are using a BioFrac fraction collector, you must program both

the BioLogic LP workstation and the BioFrac fraction collector to use them
together.

1. On the BioLogic LP system, select the appropriate fraction collector.

a.

Press the FRAC COLL button on the BioLogic LP system’s front
panel.

b.

Press the MODEL soft key.

11

4006073D.qxp 8/17/2009 7:45 AM Page 11