Specifications, Classification, Approvals – Integra LifeSciences Mirror, DenLite Illuminate User Manual

Page 10

Advertising
background image

-Page 10 of 14-

The batteries are not loaded
properly.

Reload the batteries (page 5) and
firmly press each battery down to
the bottom of the charging pocket.

Charging is complete.

Remove the batteries from the I-
C

3

Battery Charger and use.

The batteries are damaged.

Discontinue use of these batteries.

The charger green light
does not turn on or it
flashes.

The charger is outside of its
operating temperature.

Allow the charger to adjust to
room temperature and re-start the
charging process.


Specifications

Length

7.25” (18.4 cm) (without mirror and in the off/detent position)

Max Diameter

0.875” (2.2 cm)

Weight

2.50 oz (71.1 grams) (without mirror)

Operating
Conditions

60° F to 77° F
16° C to 25° C
900 hPa to 1060 hPa
50 to 95% RH

Storage
Conditions

-4° F at 50% RH to 120° F at 95% RH
-20° C at 50% RH to 50° C at 95% RH
500 hPa to 1060 hPa

Performance

45 minutes nominal On

Electrical

1.2 V NiMh batteries (qty 2 AAA)
Maximum Rating 800 mAh


Classification

Type of protection against electric shock

Internally Powered Device @ 2 x
1.2V dc 800 mA hr maximum

Degree of protection against electric shock

Type BF

Degree of protection against ingress of water

IPX3

Degree of safety of application in the presence of a
flammable anesthetic mixture with air, oxygen, nitrous oxide

Device not suitable for use in
flammable environment.

Mode of Operation

Continuous

LED Product

Class 1


Approvals

CERTIFIED TO: CAN/CSA C22.2 No. 601.1-M90,
including amendments 1 and 2; EN/IEC 60825-1; UL
60601-1

The CE mark on this product indicates that it has been
tested to and conforms with the provisions noted within
the 93/42/EEC Medical Device Directive.
EN60601-1-2:2001
EN60601-1:1990

Advertising