Led light cartridge, I-c3 battery charger, Cleaning – Integra LifeSciences Mirror, DenLite Illuminate User Manual

Page 8: Disinfection

Advertising
background image

-Page 8 of 14-

LED Light Cartridge

Caution The LED Light Cartridge may get hot during extended use. Allow the
LED Light Cartridge to cool before removing it.


If the LED Light Cartridge no longer illuminates and you have replaced the batteries,
discontinue use and replace the LED Light Cartridge.

1. Remove the collet assembly.
2. Remove the LED Light Cartridge and replace with a new LED Light Cartridge (part

number DP5150).

3. Reassemble the DenLite Illuminated Dental Mirror, page 6.


I-C3 Battery Charger

WARNING To reduce the risk of injury, charge only I-C

3

, NiMH or NiCd rechargeable

batteries. Other battery types may burst, causing personal injury or damage.


Do not leave batteries in an unplugged charger.

Cleaning

WARNING Do not re-use or sterilize the Single-Use mirrors.

Caution Do not immerse or soak the handle, collet assembly, LED light cartridge, or
batteries of the DenLite in any liquid.

You can wipe the handle, and collet assembly with cold steriliants and disinfectants.

You can wipe the LED Light Cartridge clean with a cotton swab and warm water.

You can wipe the surface of the I-C

3

Battery Charger with a damp cloth.


Disinfection

WARNING Do not re-use or sterilize the Single-Use mirrors.

Caution Do not immerse or soak the handle, collet assembly, LED light cartridge, or
batteries of the DenLite in any liquid.


Use the following solutions to disinfect the DenLite handle and collet assembly.

• Dual synergized quatennary solution (17.2% isopropanol, 0.28% Di

isobutylphenoxyethoxyethyl, dimethyl benzyl ammonium chloride) such as CaviCide

®

brand surface disinfectant

• O-phenylphenol solution (0.12%) such as Citrace

®

brand hospital germicidal deodorizer

• Isopropanol Alcohol 70%

Advertising