Guidance and manufacturer’s declarations, Page 11 of 14 – Integra LifeSciences Mirror, DenLite Illuminate User Manual

Page 11

Advertising
background image

-Page 11 of 14-


Authorized European Representative Address:
E.U. Representative

Miltex GmbH
78604 Rietheim-Weilheim/Germany

CB Scheme Certificate

IEC 6060-1-1


Guidance and Manufacturer’s Declarations

Emissions for all EQUIPMENT and SYSTEMS

The REF DP5000 DenLite Illuminated Dental Mirror is intended for use in the electromagnetic
environment specified below. The customer or user of the DP5000 DenLite Illuminated Dental Mirror
should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF Emissions
CISPR 11

Group 1

The DP5000 DenLite Illuminated Dental Mirror uses RF energy only
for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment.

RF Emissions
CISPR 11

Class B

Harmonics
IEC 61000-3-2

N/A

Flicker
IEC 61000-3-3

N/A

The DP5000 DenLite Illuminated Dental Mirror are suitable for use
in all establishments other than domestic, and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.


Immunity for all EQUIPMENT and SYSTEMS

The REF DP5000 DenLite Illuminated Dental Mirror is intended for use in the electromagnetic
environment specified below. The customer or user of the REF DP5000 DenLite Illuminated Dental
Mirror should assure that it is used in such an environment.

Immunity Test

IEC 60601 Test
Level

Compliance
Level

Electromagnetic Environment -
Guidance

ESD

IEC 61000-4-2

± 6kV Contact
± 8kV Air

± 6kV Contact
± 8kV Air

Floors should be wood, concrete, or
ceramic tile. If floors are synthetic
material, the relative humidity should
be at least 30%.

EFT

IEC 61000-4-4

±2kV Mains
±1kV I/O

N/A

Mains power quality should be that
of a typical commercial or hospital
environment.

Surge

IEC 61000-4-5

±1kV Differential
±2kV Common

N/A

Mains power quality should be that
of a typical commercial or hospital
environment.

Voltage Dips,
short interruptions,
and voltage
variations on
power supply input
lines.
IEC 61000-4-11

>95% Dip in 0.5
Cycle

60% Dip in 5
Cycles

30% Dip for 25
Cycles

>95% Dip in 5
Seconds

N/A

Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the REF
DP5000 DenLite Illuminated Dental
Mirror requires continued operation
during power mains interruptions, it
is recommended that the REF
DP5000 DenLite Illuminated Dental
Mirror be powered from an
uninterruptible power supply or
battery.

Power Frequency

50/60Hz

Magnetic Field
IEC 61000-4-8

3 A/m

N/A

Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.