Warnings

A warning statement in this manual identifies a condition or practice which, if not corrected or
discontinued immediately, could lead to patient injury, illness, or death.

WARNING Do not re-use or sterilize the Single-Use mirrors.

WARNING The LED DenLite light output is in a range that may cause premature
polymerization of composite materials.

WARNING Do not look directly at an illuminated LED Light Cartridge. Temporary
blinding may result.

WARNING Do not use the I-C

3

Battery Charger in the patient area.

WARNING To reduce risk of injury, charge only I-C

3

, NiMH, and NiCd rechargeable

batteries. Other types of batteries may burst, causing personal injury and damage.

WARNING To reduce the risk of electric shock, unplug the charger from the outlet
before cleaning. Never immerse the charger in water or any other liquid. Do not
expose the charger to rain or snow.

WARNING Never disassemble the charger or alter the plug. This may cause electric
shock.

WARNING To reduce the risk of injury, charge only I-C

3

, NiMH or NiCd rechargeable

batteries. Other types of batteries may burst, causing personal injury or damage.

WARNING To reduce risk of battery explosion, follow the instructions and cautionary
markings on the battery charger, battery, and DenLite Illuminated dental Mirror.

WARNING The batteries may become warm during charging. Use caution when
handling.

WARNING The LED Light Cartridge may get hot during extended use. Allow the
instrument to cool before removing it.

WARNING Do not use the DenLite Illuminated Dental Mirror continuously for more
than 45 minutes.


Cautions

A caution statement in this manual identifies a condition or practice which, if not corrected or
discontinued immediately, could lead to equipment failure, equipment damage, or data loss.

Caution Never attempt to charge single-use alkaline or zinc carbon batteries in the
charger.

Caution Do not charge Rechargeable Alkaline (Renewal) batteries in the battery
charger.

Caution Do not immerse or soak the handle, collet assembly, LED light cartridge, or
batteries of the DenLite in any liquid.

Caution Do not autoclave the batteries or the LED Light Cartridge. Damage to the
batteries and LED Light Cartridge will occur.