HI-6053
AND
HI-6005
F
IELD
P
ROBES
Probe
Command
Description
Probe Response
B<CR>
Read battery voltage
:Bxx.xx<CR>
• :B04.80
=100% of charge
=4.8 volts
• :B04.00
=0% of charge
=4.0 volts
BP
Read battery capacity in
percentage
Bxx<N or F><CR>
• :B64N
=100% of charge
=4.8 volts
• :B00F
=0% of charge
=4.0 volts
• N=safe operating level
• F=fail level
Percentage reported in
hexadecimal, 0% to 64%
The BP command is provided for backward compatibility and should
not be used to monitor the converter voltage, which always responds
with :B64N<CR>.