ETS-Lindgren HI-4422 Field Probe (Archived) User Manual
Page 88

3.
Press the DISCHARGE button on the battery charger. The batteries
will fully discharge, and then fully recharge. The total time required to
discharge and recharge is approximately seven hours.
S
PECIFICATIONS
B
ATTERY
C
HARGER
S
PECIFICATIONS
The battery charger may be powered by standard line voltage (110–240 VAC,
50–60 Hz) or by an optional automobile cigarette lighter plug (12.5 VDC).
Power
Main: IEC filtered AC power input module
110-240 VAC, 500 mA max,
50-100 Hz
Alternate:
Automobile cigarette lighter to 2 mm
power plug adapter cord, 12.5 Vdc,
100 mA
Fuses:
250 volt, 1.0 Amp, Type T
(5 mm x 20 mm)
Output
Open Circuit Voltage:
12 Vdc
Fast Charge Pending Current:
2.0 mA
Fast Charge Current:
220 mA
Pulsed Trickle Charge Current:
2.0 mA
Output Voltage (During Fast
Charge):
4–8 Vdc
88
|
Series H-491198-01 Battery Charger for NiMH Batteries
Archived 6/29/10
Use only the content for
HI-4422 in this manual.
Disregard content for
all other probe models;
see ETS-Lindgren
website for other
probe content.